From reading to result: an uncertainty budget

Turn repeated readings and an instrument's specification into a result with an honest uncertainty: Type A and Type B, combining, coverage and rounding, in a budget you can defend.

Lesson 1 of 350 minவரைவு

இந்தப் பாடம் மொழிபெயர்க்கப்படும் வரை ஆங்கிலத்தில் உள்ளது.

What the standard asks

Section 7 of the 2026 Electronics standard, Measurement, documentation, and reporting, is worth 15% of the marks (WorldSkills Occupational Standards, Electronics, WorldSkills Shanghai 2026, checked 2026-11-25). It asks you to know what makes a measurement good quality and how measurements and calculations are checked, and to record each measured value next to its design value.

The section page showed how to work out a meter's limit from its specification, ±(% of reading + digits). This lesson takes the next step. Most results are not one reading: they are calculated from several readings, each with its own doubt. An uncertainty budget is the table that shows where the doubt in a result comes from, and how big it is in total.

The theory

Error and uncertainty

Error is the difference between your result and the true value. You never know it exactly; if you did, you would correct it. Uncertainty is how far from your result the true value could reasonably be. A result without one cannot be compared with anything.

Some errors you can, and must, correct before you work out the uncertainty. They push every reading the same way, and are called systematic. Example: a 10 MΩ meter across the lower resistor of a 1 MΩ and 1 MΩ divider on 5.000 V. In parallel with the meter, the lower resistor becomes 909 kΩ, so the meter reads 5 × 909 ÷ 1909 = 2.381 V instead of 2.500 V: −4.8%, every time. Averaging more readings never removes it. Work out the loading and correct for it, or use a meter whose input resistance is far higher than the circuit's.

Type A: from the scatter of repeated readings

Take n readings. Their mean is your best value. The sample standard deviation, s, says how far single readings scatter. The standard uncertainty of the mean is:

u_A = s ÷ √n

Ten readings make the mean's uncertainty about a third of one reading's (√10 = 3.16). A hundred make it a tenth. More readings shrink only this term.

Type B: from everything else

Some uncertainties do not come from scatter: the meter's specification, a resistor's tolerance, the display's resolution, a calibration certificate. When a source gives only limits, ±a, and any value between them is as likely as any other, the standard uncertainty is:

u_B = a ÷ √3

This is the rectangular distribution (NIST Technical Note 1297). Resolution is a case of it: a display that moves in steps of d can hide up to ±d/2, so u = d ÷ √12. The Uno's 10-bit ADC with a 5 V reference has d = 5 V ÷ 1024 = 4.88 mV, so u = 1.41 mV from resolution alone.

Combining

  • For a sum or a difference, add the standard uncertainties in quadrature: u = √(u1² + u2² + ...).
  • For a product or a quotient, do the same with relative uncertainties. For I = V ÷ R: (u_I/I)² = (u_V/V)² + (u_R/R)².

Quadrature, not plain addition, because independent errors rarely all push the same way at once.

Coverage and rounding

The combined standard uncertainty, u_c, covers the true value with about 68% confidence. Reports usually give the expanded uncertainty U = k × u_c with k = 2, about 95%. Always say which k you used.

Round U to two significant figures, then round the result to the same decimal place. Write 3.014 mA ± 0.045 mA (k = 2). Not 3.0145454 mA ± 0.04469 mA, and not 3.0 mA ± 0.045 mA.

The budget

One row for each source: what it is, Type A or B, its limit or its s, the divisor, its standard uncertainty, and its share of the total. The share tells you where to spend your effort. Halving the smallest term changes almost nothing.

Worked example

Purpose. Find the current through an LED from the voltage across its 220 Ω ±1% series resistor, without breaking the circuit.

Readings. A meter on its 2 V range, 0.001 V a digit, specification ±(0.5% + 2 digits). Ten readings, the leads lifted and replaced each time:

0.662, 0.663, 0.663, 0.664, 0.662, 0.663, 0.665, 0.663, 0.664, 0.663 V.

  1. Mean and scatter. Mean V = 0.6632 V and s = 0.92 mV, so u_A = 0.92 ÷ √10 = 0.29 mV.
  2. The meter. a = 0.5% × 0.6632 V + 2 × 0.001 V = 3.32 mV + 2.00 mV = 5.32 mV, so u_B = 5.32 ÷ √3 = 3.07 mV.
  3. The voltage. u_V = √(0.29² + 3.07²) = 3.08 mV: 0.46% of 0.6632 V.
  4. The resistor. ±1%, rectangular: u_R/R = 1% ÷ √3 = 0.58%.
  5. The current. I = 0.6632 V ÷ 220 Ω = 3.0145 mA. u_I/I = √(0.46² + 0.58²) = 0.74%, so u_I = 0.0223 mA.
  6. Expanded. U = 2 × 0.0223 mA = 0.045 mA.

Result: I = 3.014 mA ± 0.045 mA (k = 2).

SourceTypeLimit or sDivisorStandard uncertaintyShare of the total
Scatter of 10 readingsAs = 0.92 mV√100.29 mV (0.04%)under 1%
Meter specificationB±5.32 mV√33.07 mV (0.46%)39%
Resistor toleranceB±1%√30.58%61%

The shares are of u_I², because squares are what add. They tell the story. The scatter, where most students spend their effort, is under 1% of the total. The resistor's tolerance is the largest term. To do better, measure the resistor itself with the power off, or fit a 0.1% resistor. Ten more readings would change nothing you could see.

The downloadable spreadsheet does all of this from your own readings and your own meter's specification.

Practice task

  1. Measure the voltage across one resistor in a working circuit 10 times, lifting and replacing the leads each time. Enter your readings and your meter's specification, from its manual, in uncertainty-budget.csv.
  2. Measure the resistor out of the circuit, with the specification for that range, and use its value and uncertainty instead of the tolerance. How much smaller is U now?
  3. Load adc_statistics.ino on an Uno with a potentiometer on A0. Measure the Uno's 5V pin with your meter and put the value in VREF_MV. Record the mean and s at three knob positions.
  4. Turn one ADC mean into millivolts with its uncertainty, from three terms: s ÷ √n, the resolution (4.88 mV ÷ √12), and the uncertainty of your 5V reading. Which term is largest? Compare the result with your meter on A0.
  5. Write every result as "value ± U (k = 2)", rounded correctly, with its budget table under it.

Common mistakes

  • Taking the scatter as the whole uncertainty. With a good meter, its specification is usually far bigger.
  • Adding uncertainties in a straight line. Independent terms combine in quadrature.
  • Using a ±limit as if it were a standard uncertainty. Divide by √3 first for a rectangular limit, or by the k that a calibration certificate gives.
  • Giving U without its k.
  • Mixing absolute and relative uncertainties in one sum.
  • Too many digits. An uncertainty of 0.04469 mA is not known to four figures.
  • Calling a loading error "uncertainty". It is systematic: correct it first.
  • Thinking identical readings mean no uncertainty. They only mean the scatter is smaller than one step of the display.

Self-check

  1. Ten readings have s = 1.2 mV. What is the standard uncertainty of their mean? How many readings would halve it?
  2. A meter's limit on a reading is ±6 mV. What standard uncertainty goes in the budget, and why that divisor?
  3. You calculate P = V × I. V has a relative standard uncertainty of 0.3%, and I has 0.4%. What is P's?
  4. 100 ADC readings on an Uno are all 600. Is the uncertainty of the reading zero? Which terms must you still include?
  5. A 10 MΩ meter reads 2.381 V across the lower resistor of a 1 MΩ and 1 MΩ divider on 5.000 V. Is the 0.119 V difference part of the uncertainty? What do you do with it?
  6. Write this result correctly: R = 4712.38 Ω, U = 23.418 Ω, k = 2.
  7. In the worked example, which single change cuts U the most: 20 more readings, a meter twice as good, or the resistor measured to ±0.1%?
Show answers
  1. 1.2 ÷ √10 = 0.38 mV. To halve it you need four times as many readings: 40.
  2. 6 ÷ √3 = 3.5 mV. The specification gives only limits, and any value inside them is taken as equally likely: a rectangular distribution, whose standard uncertainty is its half-width divided by √3.
  3. √(0.3² + 0.4²) = 0.5%. For a product, relative uncertainties combine in quadrature.
  4. No. Identical readings only mean the scatter is smaller than one count. Include the resolution, 4.88 mV ÷ √12 = 1.41 mV, and the uncertainty of the 5 V reference, which is usually larger still.
  5. No. It is a systematic error: the meter's 10 MΩ loads the 1 MΩ resistor, the same way every time. Correct for it (the voltage without the meter is 2.500 V), or use a meter with a much higher input resistance. Then work out the uncertainty.
  6. R = 4712 Ω ± 23 Ω (k = 2). U to two significant figures, and the result to the same place.
  7. Measuring the resistor. It is 61% of the total; the meter is 39% and the scatter under 1%. With the resistor known to ±0.1%, U falls from 0.045 mA to about 0.028 mA. A meter twice as good gives about 0.038 mA. Twenty more readings change nothing you could see.

Downloads

  • uncertainty-budget.csv: the worked example as a spreadsheet. Put your readings in B4 to B13 and your meter's specification in B20 to B22: the budget, the result and each source's share work themselves out.
  • adc_statistics.ino: an Uno takes 100 readings of A0 and prints the mean, s and s ÷ √n every 3 seconds: a Type A evaluation done by the board. Save it in a folder named adc_statistics.

Practise in ARGYNIX Lab

Open the ADC statistics project in ARGYNIX Lab (புதிய தாவலில் திறக்கும்): an Uno with a potentiometer on A0, running adc_statistics.ino. Every 3 seconds the Serial page shows the mean of 100 readings, s, s ÷ √n, and the lowest and highest reading.

Turn the knob between reports. The mean follows it, but s stays at 0.00, and the lowest equals the highest. The Lab's ADC has no noise. That is this lesson's point, in front of you: no scatter does not mean no uncertainty. With s = 0, your budget still holds the resolution, 4.88 mV ÷ √12 = 1.41 mV, and the 5 V reference, which in the Lab is exactly 5.000 V and on a real Uno is not.

Write the budget for one Lab reading, then run the same sketch on a real Uno. A real board usually shows some scatter, often a count or two, and its 5V pin is rarely exactly 5.000 V. Compare the two budgets: which term changed most?

Run it in your browser, then do it again on your own board.

Open in ARGYNIX Lab (புதிய தாவலில் திறக்கும்)

இந்தச் சாதனத்தில் மட்டுமே சேமிக்கப்படும்.

ஆதாரம்: WorldSkills Occupational Standard, Electronics, 2026; nist.gov; bipm.org; ww1.microchip.com. சரிபார்த்தது 25 நவம்பர், 2026.

அதிகாரபூர்வ தரநிலையிலிருந்து எங்கள் சொந்த வார்த்தைகளில் எழுதப்பட்டது. இது அதிகாரபூர்வ WorldSkills ஆவணம் அல்ல.