Oscilloscope measurements you can defend

Set up the probe, the sample rate and the trigger so a scope reading is right, then measure frequency, duty, amplitude and rise time, and know how far to trust each number.

Lesson 2 of 355 minDraft

What the standard asks

Section 7, Measurement, documentation, and reporting, is worth 15% of the marks (WorldSkills Occupational Standards, Electronics, WorldSkills Shanghai 2026, checked 2026-11-25). It asks you to record each measurement next to its design value, such as amplitude, frequency, duty cycle or cutoff frequency, and to complete reports with waveforms.

The section page listed those measurements and the limit a scope's bandwidth sets. This lesson is about the settings behind a reading: the probe, the sample rate, the trigger and the levels the scope measures at. Get those wrong and a number with four digits can be wrong in the first one.

The theory

The probe is part of the circuit

  • A 10x probe puts about 10 MΩ in parallel with 10 to 15 pF on the point you touch. Use it by default.
  • A 1x probe puts about 1 MΩ in parallel with its cable's capacitance, often around 100 pF, and its bandwidth is often only a few megahertz. At 1 MHz, 100 pF is a load of only 1.6 kΩ: enough to slow an edge and pull down any high-impedance node.

Compensate before you measure

A 10x probe has a small trimmer that matches it to the scope's input. Clip it to the scope's square-wave output, usually 1 kHz, and adjust until the top of the wave is flat. Rounded leading corners mean under-compensated: edges look slower than they are, and high-frequency amplitudes read low. Spikes on the corners mean over-compensated. A probe moved to another channel or another scope needs doing again. It takes 20 seconds.

The ground lead rings

The usual 15 cm ground clip makes a loop of about 150 nH (a rough guide is 1 nH per millimetre of wire). With the probe's 10 pF it resonates at 1 ÷ (2π√(150 nH × 10 pF)) ≈ 130 MHz. A fast edge rings at that frequency, and the ringing belongs to the probe, not to your circuit. For edges and rise times, use the short ground spring on the probe tip. At about 10 nH it moves the ringing to around 500 MHz, beyond a 100 MHz scope.

The sample rate falls when you zoom out

A scope keeps a fixed number of points for each screen, so its sample rate is the record length divided by the time on the screen. With 10 000 points:

  • at 100 µs/div (1 ms on the screen): 10 MS/s;
  • at 100 ms/div (1 s on the screen): 10 kS/s.

At 10 kS/s a 1 kHz PWM gets only 10 samples a period, and anything above 5 kHz is aliased: it shows as a false, slower wave. The screen looks calm and is wrong. Two rules: at least 4 to 5 samples in each period of the fastest frequency you care about, and at least 10 samples on any edge you time. Read the sample rate on the screen before you trust a waveform.

A picture that stands still

Trigger on a clean edge of the signal you measure, at about half its amplitude. For two signals from one timer, trigger on one and read the other against it. The test sketch puts 25% and 50% PWM on D9 and D10 from the same timer, so their rising edges line up.

Where the scope measures

Automatic measurements use reference levels on each edge, usually 10%, 50% and 90% of the way from the base to the top. Width and duty are timed at 50%, rise time from 10% to 90%. Amplitude is top minus base, which ignores overshoot; peak-to-peak includes it. Know which one the task asks for.

Automatic values come from what is on the screen. Show at least 3 to 5 periods, and check the important values with cursors. A cursor moves one pixel at a time: at 50 µs/div, 500 µs across a screen 1000 pixels wide, each pixel is 0.5 µs.

Rise time and the scope's own limit

A scope and its probe have their own rise time, about 0.35 ÷ bandwidth: 3.5 ns for 100 MHz. What you see is a mix of the signal's rise time and the scope's:

t_shown² ≈ t_signal² + t_scope²

So t_signal ≈ √(t_shown² − t_scope²). Trust this correction only when the shown rise time is at least about 1.5 times the scope's own. Below that, report that the edge is limited by the scope, with the value shown, not a corrected value.

How far to trust the number

  • Time and frequency. A scope's timebase is usually good to tens of parts per million. Your limits are the sample interval, the cursor's pixel and the signal's own jitter.
  • Voltage. An 8-bit scope divides its 8 vertical divisions into 256 steps: at 1 V/div, one step is 31 mV. Its DC gain accuracy is often ±2 to 3% of full scale. Take the figure from your scope's datasheet and put it in your budget, as in the uncertainty lesson. Fill the screen: a 5 V signal at 2 V/div uses only a third of the screen's height and wastes resolution.

Worked example

Purpose. Measure the 1 kHz, 25% signal on D9 of an Uno running test_signals.ino, and record it so someone else can check it.

Setup. A 100 MHz scope, 10x probe compensated, ground spring, CH1 on D9, DC coupling, 1 V/div, trigger CH1 rising at 2.5 V. For frequency, duty and amplitude: 500 µs/div, so 5 ms and 5 periods on the screen. With 10 000 points that is 2 MS/s, 0.5 µs a sample and 500 samples on each HIGH time. For the HIGH time with cursors: 50 µs/div. For the rise time: 5 ns/div.

QuantityDesignMeasuredMethodDifference
Frequency1000 Hz1002.3 Hzautomatic, 5 periods+0.23%
HIGH time250.0 µs249.5 µscursors, 50 µs/div−0.20%
Duty25.0%25.0%automatic0.0%
Amplitude, top minus base5.0 V4.94 Vautomatic−1.2%
Rise time, 10% to 90%not specified6.1 ns shownautomatic, 5 ns/divcorrected below

Reading the results.

  1. Frequency and HIGH time have one cause. The Uno's ATmega328P runs from a ceramic resonator, not a crystal, so its 16 MHz is a little off. Both readings point to the same clock: 1002.3 Hz × 16 000 counts = 16.04 MHz, and 4000 counts at 16.04 MHz last 249.4 µs. The code is right; the clock is not exact.
  2. The duty is exact because it is a ratio of two times from the same clock: the clock error cancels.
  3. The amplitude is the 5V rail less the output's small drop, and a USB supply is rarely exactly 5.00 V. Measure the 5V pin with a meter to separate the two.
  4. The rise time. 6.1 ns is 1.7 times the scope's 3.5 ns, so the correction is fair: √(6.1² − 3.5²) = √(37.2 − 12.3) = 5.0 ns. Report "5.0 ns, corrected for the scope's rise time of 3.5 ns". With the long ground clip instead of the spring, expect ringing near 130 MHz on this edge and a rise time that changes with how the lead lies: that is not a measurement.

Each row then goes into the downloadable record, with the saved waveform's file name. A reading without its settings cannot be checked, so it earns little in a report.

Practice task

  1. Compensate both probes on your scope's calibrator. Save pictures of an under-, an over- and a correctly compensated wave for your notes.
  2. Load test_signals.ino on an Uno. Fill one block of scope-record.md each for D9, D10 and D13.
  3. Measure D9's frequency at 200 µs/div, 20 ms/div and 1 s/div. Write down the sample rate at each. At which setting does the reading go wrong, and why?
  4. Look at one rising edge of D9 with the long ground clip, then with the ground spring. Save both. What frequency is the ringing? Compare it with the 1 ÷ (2π√(LC)) estimate.
  5. Check two automatic values with cursors, and write down the cursor resolution you had.
  6. From your measured frequency, work out your Uno's real clock frequency.

Common mistakes

  • Not compensating a probe after moving it to another channel or scope.
  • The long ground clip on fast edges. You measure the probe's ringing, not the circuit.
  • Never looking at the sample rate. An aliased wave looks as calm as a real one.
  • One period, or part of one, on the screen for an automatic measurement.
  • Peak-to-peak when the task asks for amplitude, or the other way round.
  • Correcting a rise time that is close to the scope's own and reporting it as a value.
  • AC coupling on a PWM signal. The DC level is lost, and the levels move as the duty changes.
  • A saved picture without its sample rate, trigger and probe setting.

Self-check

  1. Your scope keeps 10 000 points. What is its sample rate at 50 ms/div? Can it measure the duty of a 20 kHz PWM there?
  2. Why does a 10x probe load a circuit less than a 1x probe? Give typical values.
  3. A probe is under-compensated. What does a 1 kHz square wave look like, and what happens to amplitude readings at high frequency?
  4. With a 15 cm ground lead, a clean edge shows ringing near 130 MHz. Is the circuit ringing? What do you change?
  5. A 100 MHz scope shows a rise time of 4.0 ns. What do you report?
  6. D9 measures 1002.3 Hz and 25.0% duty. Why is the frequency off by 0.23% but the duty exact?
  7. A 5 V step has 0.8 V of overshoot. The scope gives Vpp = 5.8 V and amplitude = 5.0 V. The task asks for the amplitude. What do you write, and why do the two differ?
Show answers
  1. 0.5 s on the screen, so 20 kS/s: one sample in each period of 20 kHz. No: the signal is aliased and the duty means nothing. Zoom in until a few periods fill the screen: at 20 µs/div, 200 µs on the screen, the sample rate is 50 MS/s.
  2. A 10x probe puts about 10 MΩ and 10 to 15 pF on the circuit. A 1x probe puts about 1 MΩ and often around 100 pF, its cable's capacitance. At high frequency the capacitance matters most: 100 pF is only 1.6 kΩ at 1 MHz.
  3. The leading corners are rounded, and the wave creeps up to its flat top. The probe passes high frequencies too little, so edges look slower and high-frequency amplitudes read low. Adjust its trimmer on the calibrator.
  4. Probably not. The ground lead's loop and the probe's capacitance resonate near 130 MHz. Use the ground spring, as close to the point as you can, and measure again.
  5. 4.0 ns is only 1.14 times the scope's 3.5 ns, so a correction is not reliable. Report that the rise time is limited by the scope: shown as 4.0 ns, with the scope's own rise time of 3.5 ns.
  6. The frequency depends on the Uno's clock, a ceramic resonator about 0.2% fast. The duty is the ratio of two times from the same clock, so the clock error cancels.
  7. 5.0 V. Amplitude is top minus base and ignores the overshoot; peak-to-peak includes the overshoot's peak. Write which one you measured, and record the overshoot separately if it matters.

Downloads

  • scope-record.md: our record for one scope measurement: the instrument, every setting, each result beside its design value with the method used, and the rule for correcting a rise time.
  • test_signals.ino: known signals on an Uno to practise on: 1 kHz at 25% on D9 and at 50% on D10 from one timer, and a 2 Hz square wave on D13. Save it in a folder named test_signals.

Practise in ARGYNIX Lab

Open the test signals in ARGYNIX Lab (opens in a new tab): an Uno with LEDs on D9, D10 and D13, running test_signals.ino. Open the Logic page: D9 reads about 25%, D10 about 50%, and D13 switches every 250 ms.

The Logic page shows each pin's share of time HIGH, about 60 times a second. A 1 kHz signal appears as its duty, not as separate pulses. Use it to check duties and slow timing; edges, amplitude and ringing need a real scope and a real probe.

Use it to predict before you go to the bench. Change OCR1A = 3999 to 11999 and run: D9 should read 75%. Put it back, then change TOP to 7999 and leave OCR1A and OCR1B as they were. Work out the new frequency and both duties before you run, then check the duties on the Logic page. (D9 reads 50%, and D10 stays HIGH all the time: OCR1B is now equal to TOP.) Then measure the same changes on your Uno with the scope, and fill in the record.

Run it in your browser, then do it again on your own board.